Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
71
OutputFormat=4 (pressure and scan number in Hex)
Data is output in the order listed, with no spaces or commas between
parameters. Shown with each parameter is the number of digits, and how to
calculate the parameter from the data (use the decimal equivalent of the hex
data in the equations).
The 19plus V2 is automatically set to OutputFormat=4 when
communicating with:
• Auto Fire Module (AFM) used with an SBE 32 Carousel Water
Sampler - This allows the AFM to use the 19plus V2 pressure data to
determine when to close SBE 32 Carousel water bottles.
• SBE 55 ECO Water Sampler - This allows the SBE 55 to use the
19plus V2 pressure data to determine when to close its water bottles if set
up for autonomous water sampling.
1. (strain-gauge or Quartz - PType=1 or 3) pressure (decibars) = pppp - 100
2. Scan number = ssssss
OutputFormat=5 (engineering units in decimal, in XML)
Data is output in the order listed, with no carriage return or line feed between
each parameter (however, there is a carriage return and line feed at the end of
the data stream, after the </datapacket> closing tag). Shown with each
parameter is the number of digits and the placement of the decimal point.
Leading zeroes are suppressed, except for one zero to the left of the
decimal point.
: 19
V2 used with AFM and Carousel,
example scan = ppppssssss
= 00C80001F0
• Pressure = pppp = 00C8 (200 decimal);
pressure (decibars) = 200 - 100 = 100 decibars
• Scan number = ssssss = 0001F0 (496 decimal);
scan number = 496
OutputFormat=5 is listed in the
GetCD and DS response as
converted XML UVIC.
OutputFormat=4 is listed in the
GetCD and DS response as
converted HEX for AFM.