Manual revision 006 Section 2: Description of SBE 19plus V2 SBE 19plus V2
19
Data Storage
The SBE 19plus V2 has a 64 Mbyte FLASH memory. Shown below
are calculations of available data storage for several configurations.
See Specifications for storage space required for each parameter.
Data I/O
The SBE 19plus V2 receives setup instructions and outputs diagnostic
information or previously recorded data via a three-wire RS-232C link, and is
factory-configured for 9600 baud, 8 data bits, 1 stop bit, and no parity.
19plus V2 RS-232 levels are directly compatible with standard serial interface
cards (IBM Asynchronous Communications Adapter or equal). The
communications baud rate can be changed using BaudRate= (see Command
Descriptions in Section 4: Deploying and Operating SBE 19plus V2).
Magnetic Reed Switch
Profiling Mode
A magnetic switch, mounted on the conductivity cell guard, can be used to
start and stop logging in Profiling mode. Sliding the switch to the On position
wakes up the SBE 19plus V2 and starts logging. Sliding the switch to the Off
position stops logging. The switch should be Off (towards the sensor end cap)
when not logging data; i.e., during setup, diagnostics, and data extraction.
The 19plus V2 can be set up to ignore the switch position:
If IgnoreSwitch=Y: logging is started and stopped with
commands sent through the terminal program. Switch position has no
effect on logging.
If AutoRun=Y: logging is started and stopped when external power is
applied and removed. Switch position has no effect on logging.
Moored Mode
In Moored mode, the magnetic switch position has no effect on logging.
Logging is started and stopped with commands sent through the
terminal program.
Example 1: Profiling mode, no auxiliary sensors, strain-gauge pressure
T & C = 6 bytes/sample Strain-gauge P = 5 bytes/sample
Storage space 64,000,000 / (6 + 5) 5,818,000 samples
Example 2: Profiling mode, 6 external voltages, strain-gauge pressure
T & C = 6 bytes/sample Strain-gauge P = 5 bytes/sample
External voltages = 2 bytes/sample * 6 voltages = 12 bytes/sample
Storage space 64,000,000 / (6 + 5 + 12) 2,782,000 samples
Example 3: Moored mode (stores date and time), 6 external voltages, strain-gauge pressure, SBE 38 RS-232 sensor
T & C = 6 bytes/sample Strain-gauge P = 5 bytes/sample SBE 38 = 3 bytes/sample
External voltages = 2 bytes/sample * 6 voltages = 12 bytes/sample Date/Time = 4 bytes/sample
Storage space 64,000,000 / (6 + 5 + 12 + 3 + 4) 2,133,000 samples
Notes:
See Command Descriptions in
Section 4: Deploying and
Operating SBE 19plus V2.
Leave the switch in the Off
position if IgnoreSwitch=Y or
AutoRun=Y, or in Moored mode. If
the switch is On, the 19plus V2
draws an additional 15 µA from the
battery while in quiescent state.
Magnetic switch
Sensor
end cap
Note:
See Battery Endurance for
power limitations.