Sea-Bird Electronics SBE 38 Bedienungsanleitung Seite 26

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 124
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 25
Manual revision 006 Section 3: Power and Communications Test SBE 19plus V2
26
Following is a description of the menus:
Menu
Description
Equivalent Command*
File
Load command file opens selected .XML
command file, and fills Send Commands
window with commands
Unload command file closes command
file, and removes commands from Send
Commands window
Exit - Exit program.
-
Communications
Connect connect to comm port
Disconnect disconnect from
comm port
Configure Establish communication
parameters (comm port and baud rate).
Disconnect and reconnect may be useful
if 19plus V2 has stopped responding
-
Command
Abort interrupt and stop 19plus V2’s
response
Send 5 second break (not applicable to
19plus V2)
Send stop command
Set local timeSet date and time to time
sent by timekeeping software on your
computer; accuracy ± 25 msec of time
provided by computer.
Set UTC Time (Greenwich Mean Time)
Set date and time to time sent by
timekeeping software on your computer;
accuracy ± 25 msec of time provided
by computer.
(press Esc key several
times for Abort)
Stop
DateTime=
DateTime=
Capture
Capture 19plus V2 responses on screen to
file, to save real-time data or use for
diagnostics. File has .cap extension.
Click Capture again to turn off capture.
Capture status displays in Status bar.
Upload
Upload data stored in memory, in format
Sea-Bird’s data processing software can use
(raw hex). Uploaded data has .xml
extension, and is then automatically
converted to a file with .hex extension.
Before using Upload: stop logging by
sending Stop.
Several status commands
and appropriate data
upload command as
applicable to user
selection of range of data
to upload (use Upload
menu if you will be
processing data with
SBE Data Processing or
viewing archived data
with Seasave)
Tools
Diagnostics log - Keep a diagnostics log.
Convert .XML data file Using Upload
menu automatically does this conversion;
tool is available if there was a problem
with the automatic conversion.
Send scriptSend XML script to 19plus
V2. May be useful if you have a number of
19plus V2s to program with same setup.
-
*See Command Descriptions in Section 4: Deploying and Operating
SBE 19plus V2.
Seitenansicht 25