Sea-Bird Electronics SBE 38 Bedienungsanleitung Seite 27

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 124
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 26
Manual revision 006 Section 3: Power and Communications Test SBE 19plus V2
27
3. If this is the first time Seaterm232 is being used, the Serial Port
Configuration dialog box displays:
Make the desired selections, and click OK.
4. Seaterm232 tries to automatically connect to the 19plus V2. As it
connects, it sends GetHD and displays the response, which provides
factory-set data such as instrument type, serial number, and firmware
version. Seaterm232 also fills the Send Commands window with
the correct list of commands for your 19plus V2.
If there is no communication:
A. In the Communications menu, select Configure. The Serial Port
Configuration dialog box appears. Select the Comm port and baud
rate for communication, and click OK. Note that the factory-set baud
rate is documented on the Configuration Sheet.
B. In the Communications menu, select Connect (if Connect is grayed
out, select Disconnect and reconnect). Seaterm232 will attempt to
connect at the baud specified in Step A, but if unsuccessful will then
cycle through all other available baud rates.
C. If there is still no communication, check cabling between the
computer and 19plus V2, and try to connect again.
D. If there is still no communication, repeat Step A with a different
comm port, and try to connect again.
After Seaterm232 displays the GetHD response, it provides an S> prompt
to indicate it is ready for the next command.
Note:
If OutputExecutedTag=Y, the 19plus
V2 does not provide an S> prompt
after the <Executed/> tag at the
end of a command response.
Note:
Seaterm232’s baud rate must be the
same as the 19plus V2 baud rate (set
with BaudRate=). Baud is factory-set
to 9600, but can be changed by the
user (see Command Descriptions in
Section 4: Deploying and
Operating SBE 19plus V2). Other
communication parameters
8 data bits, 1 stop bit, and no parity
cannot be changed.
Computer COM port and baud
rate for communication between
computer and 19plus V2.
Seaterm232 tries to connect at
this baud rate, but if
unsuccessful will cycle through
all available baud rates.
Seitenansicht 26