Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
68
OutputFormat=1 (engineering units in Hex)
Data is output in the order listed, with no spaces or commas between the
parameters. Shown with each parameter is the number of digits, and how to
calculate the parameter from the data (use the decimal equivalent of the hex
data in the equations).
1. Temperature (°C, ITS-90) = (tttttt / 100,000) - 10
2. Conductivity (S/m) = (cccccc / 1,000,000) - 1
3. (strain-gauge or Quartz - PType=1 or 3)
pressure (decibars) = (pppppp / 1,000) - 100
4. (if Volt0=Y) External voltage 0= vvvv / 13,107
5. (if Volt1=Y) External voltage 1 = vvvv / 13,107
6. (if Volt2=Y) External voltage 2 = vvvv / 13,107
7. (if Volt3=Y) External voltage 3 = vvvv / 13,107
8. (if Volt4=Y) External voltage 4 = vvvv / 13,107
9. (if Volt5=Y) External voltage 5 = vvvv / 13,107
10. (if SBE38=Y)
SBE 38 temperature (°C, ITS-90) = (tttttt / 100,000) 10
11. (if WetLabs=Y) WET Labs RS-232 sensor =wwwwxxxxyyyy
where wwww, xxxx, and yyyy are raw signal counts for each sensor;
yyyy all 0’s for dual sensor; xxxx and yyyy all 0’s for single sensor
12. (if GTD=Y or DualGTD=Y)
GTD #1 pressure (millibars) = pppppppp / 100,000
13. (if GTD=Y or DualGTD=Y)
GTD #1 temperature (°C, ITS-90) = (tttttt / 100,000) 10
14. (if DualGTD=Y)
GTD #2 pressure (millibars) = pppppppp / 100,000
15. (if DualGTD=Y)
GTD #2 temperature (°C, ITS-90) = (tttttt / 100,000) 10
16. (if Optode=Y) Optode oxygen (micromoles/liter) = (oooooo / 10,000) 10
17. Time (Moored mode [MM] only)
seconds since January 1, 2000 = ssssssss
Example: Profiling mode, 2 external voltages sampled,
example scan = ttttttccccccppppppvvvvvvvv
= 3385C40F42FE0186DE03050594
Temperature = tttttt = 3385C4 (3376580 decimal);
temperature (°C, ITS-90) = (3376580 / 100,000) - 10 = 23.7658
Conductivity = cccccc = 0F42FE (1000190 decimal);
conductivity (S/m) = (1000190 / 1,000,000) - 1 = 0.00019
Pressure = pppppp = 0186DE (100062 decimal);
pressure (decibars) = (100062 / 1,000) - 100 = 0.062
First external voltage = vvvv = 0305 (773 decimal);
voltage = 773 / 13,107 = 0.0590 volts
Second external voltage = vvvv = 0594 (1428 decimal);
voltage = 1428 / 13,107 = 0.1089 volts