Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
66
Data Output Formats
The SBE 19plus V2 stores data in a compact machine code. Data is converted
and output in the user-selected format without affecting data in memory.
Because memory data remains intact until deliberately overwritten, you can
upload in one format, then choose another format and upload again.
Output format is dependent on OutputFormat= (0, 1, 2, 3, 4, or 5) and on the
command used to retrieve the data, as detailed below. The inclusion of some
data is dependent on the system configuration - if the system does not include
the specified sensor, the corresponding data is not included in the output data
stream, shortening the data string. RS-232 sensors always output data in the
same format, regardless of OutputFormat=:
• SBE 38, GTD, and Optode data is always output in engineering units.
• WET Labs RS-232 sensor data is always output in raw counts.
If in Moored mode and outputting real-time data (MooredTxRealTime=Y)
while logging, a # sign precedes the real-time output for each sample.
Moored mode - For date and time
output, time is the time at the start of
the sample, after:
• a small amount of time
(1 to 2 seconds) for the 19plus V2 to
wake up and prepare to sample, and
• any programmed
DelayBeforeSampling=.
For example, if the 19plus V2 is
programmed to wake up and
sample at 12:00:00, and
DelayBeforeSampling=20, the output
time for the first sample will be
12:00:21 or 12:00:22.