Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
35
Moored Mode Pump Setting Recommendations
Sea-Bird provides the following recommendations for pump settings. Note that
longer pump times increase power usage, reducing battery endurance.
See Battery Endurance in Section 2: Description of SBE 19plus V2 for sample
battery endurance calculations.
Pump through Conductivity Cell Only
For most deployments, set MooredPumpMode=1 and
DelayBeforeSampling=0. The pump operates for 0.5 seconds before the
conductivity measurement is made, providing enough time to ventilate the cell
and bring in a new sample of water.
If the 19plus V2 is moored in an area with large thermal gradients, it may be
necessary to pump for a longer period of time, to eliminate any cell thermal
mass effects on the measurement. In this case, set MooredPumpMode=2 and
set DelayBeforeSampling= to a non-zero value, providing additional
ventilation time (allowing the conductivity cell temperature to equilibrate to
the water temperature) before taking the measurement.
Pump through Conductivity Cell and SBE 43 Dissolved Oxygen Sensor
Set MooredPumpMode=2.
As the pump brings new water into the SBE 43 plenum, some time is required
for the sensor to equilibrate to the new oxygen level. The time required is
dependent on the sensor’s membrane thickness, and on the water temperature.
Prior to 2007, all SBE 43s were sold with a 0.5 mil thick membrane.
Beginning in 2007, Sea-Bird began offering two membrane thicknesses
0.5 mil (faster response, typically for profiling applications) and
1.0 mil (slower response but more durable, typically for moored applications).
For a 0.5 mil thick membrane Recommended DelayBeforeSampling=
varies in a non-linear fashion, from 15 seconds at 15 °C to 30 seconds at
0 °C.
For a 1.0 mil thick membrane Recommended DelayBeforeSampling=
varies in a non-linear fashion, from 25 seconds at 15 °C to 40 seconds at
0 °C.
Pump through Conductivity Cell and Beckman- or YSI-type
Dissolved Oxygen Sensor
Set MooredPumpMode=2. Set DelayBeforeSampling= to
120 to 180 seconds, allowing time for the oxygen sensor to polarize before
taking the measurement.
Note:
See Application Note 64: SBE 43
Dissolved Oxygen Sensor
Background Information, Deployment
Recommendations, and Cleaning and
Storage for the response time curves.