Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
30
Section 4: Deploying and Operating
SBE 19plus V2
This section includes discussions of:
• Sampling modes (Profiling and Moored), including example sets
of commands
• Pump operation
• Real-time setup
• Timeout description
• Command descriptions
• Data output formats
• Optimizing data quality for Profiling applications
• Installing anti-foul fittings for Moored applications
• Deployment
• Acquiring real-time data with Seasave
• Recovery - physical handling and uploading data
• Processing data
Sampling Modes
The SBE 19plus V2 has two sampling modes for obtaining data:
• Profiling mode
• Moored mode
Descriptions and examples of the sampling modes follow. Note that the
19plus V2 response to each command is not shown in the examples. Review
the operation of the sampling modes and the commands described in
Command Descriptions before setting up your system.
Separate software manuals on
CD-ROM and Help files contain
detailed information on installation,
setup, and use of Sea-Bird’s real-
time data acquisition software and
data processing software.
• The 19plus V2 automatically enters
quiescent state after 2 minutes
without receiving a command.
•
Set OutputFormat=0 if you will be
using Sea-Bird’s real-time data
acquisition software (Seasave)
to view real-time data.