Appendix III: Command Summary SBE 45
45
Appendix III: Command Summary
CATEGORY COMMAND DESCRIPTION
Status DS
Display status.
Baud=x
x= baud rate (1200, 2400, 4800, 9600, 19200,
or 38400). Default 4800.
OutputFormat=x
x=0 (default): Output order is temperature,
conductivity, salinity, sound velocity.
x=1: Suppress space before conductivity output.
x=2: Reverse conductivity and salinity order (not for
use with Seasave, SBE Data Processing, or optional
Interface Box).
OutputCond=x
x=Y (default): Output conductivity (S/m) with data.
x=N: Do not.
OutputSal=x
x=Y: Output salinity (psu) with data.
x=N (default): Do not.
OutputSV=x
x=Y: Output sound velocity (m/sec) with data,
calculated as defined by SVAlgorithm=.
x=N (default): Do not.
SVAlgorithm=x
x=C (default): Calculate sound velocity as Chen
and Millero.
x=W: Calculate sound velocity as Wilson.
NCycles=x
x = A/D cycles to average per sample (default=4).
Setup
QS
Quit session and place MicroTSG in quiescent (sleep)
state. Sampling stops. Applicable only if PCB J1
jumper in Normal position.
Interval=x
x= interval (seconds) between samples (maximum
32767). Minimum time between samples determined
by NCycles=, desired parameters (salinity, etc.), and
baud rate.
AutoOff=x
(Functional only if J1 jumper in Normal position)
x=Y: Go to sleep (enter quiescent state) if 2 minutes
have elapsed without receiving a command or without
sampling data.
x=N: Do not automatically go to sleep.
J1 jumper - Normal
AutoRun=N
SingleSample=
Y or N
Wake up when Connect on Toolbar clicked or Enter
key pressed while in quiescent (sleep) state, wait for a
command.
J1 jumper - Normal
AutoRun=Y
SingleSample=N
Wake up when Connect on Toolbar clicked or Enter
key pressed while in quiescent (sleep) state, sample at
rate specified by Interval=. To stop sampling and get
S> prompt, type Stop and press Enter key.
J1 jumper - Normal
AutoRun=Y
SingleSample=Y
Wake up when Connect on Toolbar clicked or Enter
key pressed while in quiescent (sleep) state, take and
output a single sample, automatically go to sleep. To
wake up and get S> prompt, type Stop and press
Enter key.
J1 jumper - Autopower
AutoRun=N
SingleSample=Y or N
Wake up when power applied, wait for a command.
J1 jumper - Autopower
AutoRun=Y
SingleSample=N
Wake up when power applied, sample at rate specified
by Interval= until power removed. Required settings
for running MicroTSG in 3-wire (power, ground,
transmit) configuration or with Interface Box.
J1 jumper - Autopower
AutoRun=Y
SingleSample=Y
Wake up when power applied, take and output a
single sample. Wait for another command until
power removed.
Go
Start sampling, as defined by SingleSample=
and Interval=. Applicable if AutoRun=N, or
AutoRun=Y and you previously sent Stop to
stop sampling.
Operating
PCB’s J1
jumper interacts
with these
commands:
• Autopower –
pins 1 and 2
• Normal –
pins 2 and 3
Stop
Stop sampling.
50