Section 4: Setting Up MicroTSG SBE 45
31
Calibration Coefficients Command
DC Display calibration coefficients.
Equivalent to Coefficients on Toolbar.
The individual Coefficient Commands listed below are used to modify a
particular coefficient or date:
Temperature
TCalDate=S
S=Temperature calibration date.
TA0=F
F=Temperature A0.
TA1=F
F=Temperature A1.
TA2=F
F=Temperature A2.
TA3=F
F=Temperature A3.
Conductivity
CCalDate=S
S=Conductivity calibration date.
CG=F
F=Conductivity G.
CH=F
F=Conductivity H.
CI=F
F=Conductivity I.
CJ=F
F=Conductivity J.
WBOTC=F
F=Conductivity wbotc.
CTCor=F
F=Conductivity ctcor.
CPCor=F
F=Conductivity cpcor.
Example: (user input in bold)
S>DC
SBE45 V 1.1b 0011
temperature: 08-apr-96
TA0 = -9.420702e-05
TA1 = 2.937924e-04
TA2 = -3.739471e-06
TA3 = 1.909551e-07
conductivity: 09-apr-96
G = -1.036689e+00
H = 1.444342e-01
I = -3.112137e-04
J = 3.005941e-05
CPCOR = -9.570001e-08
CTCOR = 3.250000e-06
WBOTC = 1.968100e-05
Notes:
• See individual Coefficient
Commands below for definitions
of the data in the example.
• Dates shown are when
calibrations were performed.
Calibration coefficients are
initially factory-set and
should agree with Calibration
Certificates shipped with
MicroTSGs.
Note:
F = floating point number
S = string with no spaces