Section 4: Setting Up MicroTSG SBE 45
32
Data Output Format
Each scan ends with a carriage return <CR> and line feed <LF>. Leading
zeros are suppressed, except for one zero to the left of the decimal point.
All data is separated with a comma and a space, except as noted.
The converted output data format is:
• If OutputFormat=0 (default):
ttt.tttt, cc.ccccc, sss.ssss, vvvv.vvv
• If OutputFormat=1:
ttt.tttt,cc.ccccc, sss.ssss, vvvv.vvv
• If OutputFormat=2:
ttt.tttt, sss.ssss, cc.ccccc, vvvv.vvv
Note: OutputFormat=2 is not compatible with Seasave, SBE Data
Processing, or the optional Interface Box.
where
t = temperature (degrees Celsius, ITS-90)
c = conductivity (S/m); data sent only if OutputCond=Y
s = salinity (psu); data sent only if OutputSal=Y
v = sound velocity (meters/second), calculated by Chen-Millero
(if SVAlgorithm=C) or Wilson (if SVAlgorithm=W) equation;
data sent only if OutputSV=Y
Example 1: Sample data output when OutputFormat=0,
OutputCond=Y, OutputSal=N, and OutputSV=N:
23.7658, 0.00019
(temperature, conductivity)
Example 2: Sample data output when OutputFormat=1,
OutputCond=Y, OutputSal=N, and OutputSV=N:
23.7658,0.00019
(temperature,conductivity)
Notes:
• For OutputFormat=1, the space
before the conductivity output is
eliminated. The remainder of the
data format is unchanged from
OutputFormat=0.
• Seasave, SBE Data Processing,
and the optional Interface Box
are all compatible with any
combination of output
parameters, if OutputFormat=0
or 1.